Short protocol Homogeniser Precellys
®
24 / 24-Dual
PEQLAB_v0314_E
19
Combinable parameters/Refractory period
For safe operation the system is allowing the following combinations of program parameters.
Refractory periods are rest times in which a consecutive run cannot be started. They prevent the system
from overheating. However, the risk of overheating only exists during extensive high throughput appli-
cations using Precellys
®
24. In addition system temperature is always monitored by a temperature sensor
which will shut off the motor if the temperature rises above the limit. Therefore, from firmware version
4.1 (12/2011) on refractory periods do not exist any more in Precellys
®
24-Dual and can be skipped
by pressing 'cancel' in Precellys
®
24.
The actual firmware version of the system is displayed when pressing 'cancel' for 10 sec after switching
on the unit.
Speed [rpm]
Repetition of an agitation
interval
Duration of an inter-
val [s]
Pause between
intervals [s]
Refractory period*
2500
1 – 3
5 – 90
5 – 120
2 min
4000 - 6100
1 – 3
5 – 90
5 – 120
2 min
6200
1 – 3
5 – 90
5 – 120
5 min
5 – 30
5 – 120
6300 - 6500
1 – 3
31 – 60
15 – 120
5 min
6600 – 6800
*
1 – 3
5 – 30
30 – 120
5 min
* Precellys
®
24 only
Instructions for the use of 7 ml tubes in the Precellys
®
24-Dual
To prevent tube breakage caused by thermal expansion please comply to the following program pa-
rameters if using the 7 ml tubes in the Precellys
®
24-Dual.
Speed [rpm]
Repetition of an agitation
interval
Duration of an interval
[s]
2500
1 – 3
5 – 60
4000 - 5500
1 – 3
5 – 40
5600 - 6000
1 – 3
5 – 30
6100 - 6500
1 – 2