Page 33
Deioniser Column Replacement
Precision Hydrogen 1200cc
The deioniser column maintains the water quality of the water as it is circulated through the
system and to ensure the optimum health of the generator, the resin in the deioniser column
should be replaced every 12 months. Only Peak Scientific approved components should be used
with the Precision Hydrogen (Trace) 1200cc, please refer to page 31 for details on ordering the
required service kits.
Listed below are the instructions for replacing the deioniser column. The user will be prompted
to do this by the HMI screen displaying ‘Change DI Column’. However, in the event of a change
being required out with the 6 month schedule, this screen can also be accessed manually.
1. From the main screen pressing MENU will bring up a menu screen with an option
called COLUMN, this option should now be selected. If this is a scheduled change,
the screen will already display CHANGE D.I. COLUMN. The user should select the
CONTINUE option.
2. The two screens below will now be shown for Manual and Scheduled changes
respectively. The CHANGE button on both screens should now be selected.
3. Remove the fascia by first pulling it forwards and then sliding it up ensuring that the
HMI screen is disconnected from the unit.