Page 30
H
2
Detector Option
These are not standard serial ports, damage to serial equipment is
highly likely if connected. 24Vdc signal level exists on certain pins with
a maximum power rating of 5W.
The Precision Hydrogen 1200cc models can be purchased along with a H2 detector unit.
The detector is intended for use with a GC, it is recommended that the detector unit is
placed above the height of the GC outlet. The detector can be wall mounted.
This unit is plugged, from the ‘In’ port at the rear on the detector, into the ‘H2 Detector’
port at the rear of the generator, as shown below. This connection will provide the detec-
tor unit with power and allows communication between the generator and the detector
unit.
The detector will take samples continuously and both the detector and the Precision Hy-
drogen Trace will alarm if dangerous levels of Hydrogen are detected, the Precision Hydro-
gen Trace will then shut down.
Multiple detectors can be connected to one generator, up to 4 in number. If this is desired,
connect the first detectors ‘Out’ port to the second detectors ‘In’ port, as illustrated below.
Contact your local Peak Sales rep or distributor for ordering information on the H2 Detector.
Precision Hydrogen (Trace)
1200cc
H
2
Detector
H
2
Detector
Port
IN
Port
OUT
Port
CAUTION