Page 25
If the generator cannot load enough water to switch the sensors in the water bottle the
generator will alarm providing the user with the screen below.
The external water supply should be checked prior to pushing the recovery button
which will then re-engage the load pump and will allow the generator to continue to fill
with water.
When the system detects a low water level it starts a 3 hour timer, should the system not
receive enough water to switch the low level sensor it will then stop the generator from
producing hydrogen. This happens to ensure that your Precision Hydrogen Generator
remains in its optimum operating condition.
If the generator has reached this point the screen below will be shown on the HMI.