Page 13
Technical Specification
Precision Hydrogen
Environment
1200cc
1200cc Trace
Minimum Operating Ambient Temperature
10°C (50°F)
Maximum Operating Ambient Temperature
35°C (95°F)
Maximum Altitude
2000 m
Maximum Relative Humidity
90% Non-Condensing
Minimum Storage Temperature*
10°C (50°F)
Maximum Storage Temperature*
25°C (77°F)
Generator Outlets
Max Flow Rate
1200 cc/min
Purity
99.9995%
99.9999%
Gas outlet
1 x 1/8” Swagelok compression fitting
Delivery pressure
0-100 psi (0-6.9 bar)
Electrical Requirements
Voltage
100-230VAC
Frequency
50/60 Hz
Current
10A Max
Input Connection
C14 Connection
Power Cord (Supplied)
C13 socket to local connection (10A minimum)
Pollution Degree
2
Insulation Category
I
General
Generator Dimensions cm (W x D x H)
38.0 (14.9”) x 54.0 (21.2”) x 40.6 (15.9”)
Generator Weight
31.5Kg (69.5 lbs)
Shipping Crate Dimensions cm (W x D x H)
54.0 (21.2”) x 70.0 (27.5”) x 49.0 (19.2”)
Shipping Weight
38 kg (83.8 lbs)
Noise Level @ 1 m
20 dB
Heat Output
1000 BTU / Hr
Water Purity Requirement
<1.0µ Siemens/cm OR >1 Mohm-cm
* When taken out of storage the Generator should be allowed to acclimatize at room
temperature for a minimum of 3 hours before operation.