Page 20
Installation
Location
The generator should be located on a flat, level surface for operation - this will ensure
that the water level sensors within the unit operate correctly. Some components in the
generator produce heat during operation and therefore an adequate airflow must be
maintained around the generator to allow the ventilation system to perform efficiently.
If the generator is stored in an enclosed space the environment must be controlled via
an air conditioner or extraction fan. The provision must be made to allow the volume of
air in the room to be changed 5 times per hour. The rear of the generator will become
warm to the touch during operation - a minimum clearance of 15 cm (6”) from other
bodies is recommended.
Do not obstruct or connect the vents to any application. Failing to observe this could
lead to permanent damage of the generator.
Please ensure generator is positioned so as to permit easy disconnection if required.
Do not expose the unit to freezing conditions. Please adhere to the operating
temperatures below:
Minimum Operating Ambient Temperature:
10 °C (50 °F)
Maximum Operating Ambient Temperature:
35 °C (95 °F)
Note: It is recommended that the Precision Hydrogen 1200cc unit is left to run for 3
hours, and the Precision Hydrogen Trace 1200cc unit is left to run for 24 hours before
use.