Copyright © Peak Scientific Instruments Ltd 2021 — Precision Hydrogen 1200cc User Manual  Rev 8  RSID 2379 EN —31/08/2021

Содержание Precision Hydrogen 1200cc

Страница 1: ...Precision Hydrogen 1200cc All Models User Manual Copyright Peak Scientific Instruments Ltd 2021 Precision Hydrogen 1200cc User Manual Rev 8 RSID 2379 EN 31 08 2021...

Страница 2: ...or Safety Features 15 Unpacking 16 Fittings Kit Contents 17 Unpacking Instructions 17 Transit Plug Removal 18 Installation 19 Location 19 Generator Overview 20 General Dimensions 20 Unit Controls 21 R...

Страница 3: ...31 08 2021 How to use this Manual This manual is intended for end users and has been written so that it can either be read as a step by step guide to installation and usage or as a reference document...

Страница 4: ...ure Stackable with other Precision units Purities of up to 99 9995 Standard and 99 9999 Trace The Precision Hydrogen 1200cc is based on proven technology it produces hydrogen by electrolysis using a P...

Страница 5: ...Temperature Over 220 C Vapour Pressure 25 C 17mm CFH 20 C Water Viscosity 25 C N A Solubility in Water Negligible Specific Gravity 25 C 1 1 1 3 pH 25 C 7 9 5 in aqueous slurry Eye Protection Safety Gl...

Страница 6: ...tions If irritation persists move patient to fresh air Ingestion If resin has been confined to the mouth administer 250 ml of water as a mouthwash Avoid swallowing This product is considered non hazar...

Страница 7: ...reof the Customer must 1 Reasonably satisfy the Company that the Goods have been properly installed commissioned stored serviced and used and without prejudice to the generality of the foregoing that...

Страница 8: ...ions are fully understood or met A CAUTION notice denotes a hazard It calls attention to an operating procedure process or similar which if not correctly performed or adhered to could cause damage to...

Страница 9: ...fe of the generator and render the 3 year cell warranty void 3 Do not place the external water bottle above the height of the generator The bottle may be placed at the same level or below the generato...

Страница 10: ...2014 35 EU EN 61010 1 2010 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use Electromagnetic Compatibility Directive 2014 30 EU EN 61326 1 2013 Electrical Equipm...

Страница 11: ...ectrical Equipment Safety Regulations 2016 SI 2016 1101 as amended BS61010 1 2010 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use The Electromagnetic Compatibil...

Страница 12: ...Guidance Notes PDF produced by the Department for Business Innovation and skills for the UK and here for Europe All PEAK products that are subject to the WEEE directive are compliant with the WEEE mar...

Страница 13: ...6 9 bar Electrical Requirements Voltage 100 230VAC Frequency 50 60 Hz Current 10A Max Input Connection C14 Connection Power Cord Supplied C13 socket to local connection 10A minimum Pollution Degree 2...

Страница 14: ...uit These protons and electrons then recombine at the cathode of the electrolyser again with the aid of a precious metal catalyst to produce the hydrogen gas The high purity of the hydrogen is assured...

Страница 15: ...350cc at a pressure of 6 9bar 100psi The unit also meets the European Electromagnetic Compatibility and Low Voltage Directives The Generator is CE approved and has the CE mark attached All of the Pre...

Страница 16: ...ajor leak the generator will automatically shut down within 20 minutes 4 The hydrogen pressure inside the generator should only reach a maximum of 7 5 bar 110psi and this is controlled in two ways a G...

Страница 17: ...r the Peak Partner from where the unit was purchased Follow the unpacking instructions posted on the side of the crate It will require two people to remove the unit from the shipping crate and to mano...

Страница 18: ...udes additional Barbed Hose Fitting All of the generator output ports are located on the rear of the generator Unpacking Instructions The generator weight constitutes a two person lift and as such saf...

Страница 19: ...cess These plugs are fitted to the H2 vent and O2 vent of all hydrogen units and ensure no water from the internal water bottle can escape during transit Plugs should be removed during the installatio...

Страница 20: ...he room to be changed 5 times per hour The rear of the generator will become warm to the touch during operation a minimum clearance of 15 cm 6 from other bodies is recommended Do not obstruct or conne...

Страница 21: ...21 Generator Overview General Dimensions The Generator must always be placed on a flat level surface Failure to do so will affect the performance of the Generator 380 mm 14 9 540 mm 21 2 406mm 15 9 WA...

Страница 22: ...Master Connection Slave Connection Power Inlet Power Switch H2 Detector Connection Water Supply Fill H2 Exhaust H2 Delivery Drain Refer to page 29 for information on using Master and Slave connections...

Страница 23: ...lastic tubing can give off organic impurities which can cause ghost peaks and baseline instability Furthermore Peak Scientific recommend that compression fittings are used and that the fittings are fr...

Страница 24: ...tic Filling The Precision Hydrogen generator uses an automatic water filling system whereby sensors in the unit s internal water tanks detect when the water level reaches a minimum This then refills t...

Страница 25: ...age the load pump and will allow the generator to continue to fill with water When the system detects a low water level it starts a 3 hour timer should the system not receive enough water to switch th...

Страница 26: ...power on the generator Once the low level water sensor disengages the generator will then begin producing hydrogen as before Four and Eight litre water storage bottles can be purchased from Peak Scie...

Страница 27: ...Page 27 Water Usage The water consumption rate of the Precision Hydrogen 1200cc and Hydrogen Trace 1200cc Generator is shown in the below graph...

Страница 28: ...UNIT MUST BE EARTHED Before connecting the unit to the mains supply please check the information on the serial plate The mains supply must be of the stated AC voltage and frequency EARTH GROUND E Gree...

Страница 29: ...to the home screen where the Start option will allow the flow of hydrogen to the application Upon selection the unit will take a few moments to adjust to the desired delivery pressure If the unit has...

Страница 30: ...the rear of the generator as shown below This connection will provide the detec tor unit with power and allows communication between the generator and the detector unit The detector will take samples...

Страница 31: ...gether via the Master port on the first Precision Hydrogen Trace 1200cc and the Slave port at the rear of the next Precision Hydrogen Trace 1200cc unit The first Precision Hydrogen Trace 1200cc unit i...

Страница 32: ...of the generator can be switched off Cleaning Clean the outside of the Generator only using warm soapy water and a clean damp cloth Ensure the cloth is thoroughly rung out to remove excess fluid prio...

Страница 33: ...user will be prompted to do this by the HMI screen displaying Change DI Column However in the event of a change being required out with the 6 month schedule this screen can also be accessed manually 1...

Страница 34: ...ottom of the column 6 Remove the used column by pulling it towards you 7 Fit a new deioniser column 05 7062 into the generator and reconnect the tubing 8 Refasten the Velcro straps and re attach the f...

Страница 35: ...irements Service Schedule Purchase Interval Component Visit 12 Months Precision Hydrogen 1200cc Annual Maintenance Kit www peakscientific com ordering 24 Months Precision Hydrogen 1200cc 24 Month Inte...

Страница 36: ...t your service provider Capacity Excess Alarm Check external pipework for leaks Contact your service provider Water Level Fault Contact your service provider Low Water Supply Alarm Check water supply...

Страница 37: ...k Scientific product from the date of installation Once registered the warranty will be honoured for a period of 12 months If you wish to defer the installation of your generator you must notify Peak...

Страница 38: ...Page 38...

Страница 39: ...Page 39...

Страница 40: ...rvice to our customers With Peak Protected your laboratory s productivity becomes our top priority To discuss Peak Protected generator cover and payment options speak to your local Peak Representative...