16
WARNING
Servicing and/or repair of the Generator should only be undertaken by a
TECHNICALLY COMPETENT PERSON, with the generator in its safely isolated
condition.
SAFELY ISOLATED CONDITION
Definition: The unit is in a Safely Isolated Condition when it is disconnected from its
application, fully de-pressurised and isolated from the Electrical Supply. Directions for
isolating the generator are shown below.
Isolating the Generator:
a) Switch off the unit.
b) Unplug the generator from the mains supply and remove the power cord from the
rear of the unit.
c) Ensure the pressure gauge on the front panel reads zero. (If gauge does not fall to
zero, loosen outlet fitting slightly to allow trapped gas to escape.)
d) Disconnect from the application.
e) Important Note: Allow 10 minutes for unit to fully de-pressurise. After this time pull
out the ‘Quick Release Ring’ on the Pressure Relief valve situated on the tank.
Ensure tank is fully depressurised before releasing the ring.
SERVICE INTERVAL
COMPONENT
PART NO.
ELIMINISER FILTER (x1)
02-4366
COALESCER FILTER (x1)
02-4364
FLOAT ASSEMBLY VALVE (x2)
02-4464
RAC Filter (x1)
00-4425
ANNUAL SERVICE KIT
INCLUDING ABOVE
(08-4714)
SOLENOID VALVE 2/2 G1/8" N.C 2.4mm (x1)
02-4692
SOLENOID VALVE 2/2 G1/8M N.O 1.6mm (x2)
02-6000
2 YEARS
COMPRESSOR ASSEMBLY (x4)
08-8069
ALTERNATIVE TO COMPRESSOR
ASSEMBLY
COMPRESSOR RE-FIT (x4)*
06-5529
On Request
02-1054
1 YEAR
1 YEAR OR ON SYSTEM ALARM
INLET FILTER (x1)
Routine Maintenance
Maintenance Schedule
*Compressors can be re-fitted as an alternative to replacement up to a maximum of 3
times, this is a more cost effective solution, however a degree of technical expertise
is required and can be time consuming.
Maintenance
Nitrogen N*10DR Nitrogen Generator
UM – N*10DR – Rev 4