16

WARNING

Servicing  and/or  repair  of  the  Generator  should  only  be  undertaken  by  a 

TECHNICALLY COMPETENT PERSON, with the generator in its safely isolated 

condition.

SAFELY ISOLATED CONDITION

Definition: The unit is in a Safely Isolated Condition when it is disconnected from its 

application, fully de-pressurised and isolated from the Electrical Supply. Directions for 

isolating the generator are shown below.

Isolating the Generator:

a) Switch off the unit.

b) Unplug the generator from the mains supply and remove the power cord from the 

rear of the unit.

c) Ensure the pressure gauge on the front panel reads zero. (If gauge does not fall to 

zero, loosen outlet fitting slightly to allow trapped gas to escape.)

d) Disconnect from the application.

e) Important Note: Allow 10 minutes for unit to fully de-pressurise. After this time pull      

out  the  ‘Quick  Release  Ring’ on  the  Pressure  Relief  valve  situated  on  the  tank. 

Ensure tank is fully depressurised  before releasing the ring.

SERVICE INTERVAL

COMPONENT

PART NO.

ELIMINISER FILTER (x1)

02-4366

COALESCER FILTER  (x1)

02-4364

FLOAT ASSEMBLY VALVE (x2)

02-4464

RAC Filter (x1)

00-4425

ANNUAL SERVICE KIT

INCLUDING ABOVE

(08-4714)

SOLENOID VALVE 2/2 G1/8" N.C 2.4mm  (x1)

02-4692

SOLENOID VALVE 2/2 G1/8M  N.O 1.6mm  (x2)

02-6000

2 YEARS

COMPRESSOR ASSEMBLY (x4)

08-8069

ALTERNATIVE TO COMPRESSOR 

ASSEMBLY

COMPRESSOR RE-FIT (x4)*

06-5529

On Request

02-1054

1 YEAR

1 YEAR OR ON SYSTEM ALARM

INLET FILTER  (x1)

Routine Maintenance

Maintenance Schedule

*Compressors can be re-fitted as an alternative to replacement up to a maximum of 3 

times, this is a more cost effective solution, however a degree of technical expertise 

is required and can be time consuming.

Maintenance

Nitrogen N*10DR Nitrogen Generator

UM  – N*10DR – Rev 4

Содержание N 10DR Series

Страница 1: ...1 User Manual UM N 10DR Rev 4 Nitrogen N 10DR Nitrogen Generator...

Страница 2: ...duction 7 Unpacking and Installation 8 Removal of Transit Bracket 9 Useful Information 10 Electrical Connection 10 Generator Environment 11 General Dimensions 11 Generator Connections 12 Generator Con...

Страница 3: ...elease G L 16 10 06 2 Notes added for Indoor use only Altitude T G 10 09 07 3 Update USA address S K 25 08 2008 4 Updated solenoid valve part S K 26 01 2009 Document Change History Nitrogen N 10DR Nit...

Страница 4: ...5 Subject to the Company being notified of any defect as is referred to in sub clause 2 hereof within a reasonable time of it becoming apparent and subject always to the terms of sub clause 4 hereof...

Страница 5: ...5 Declaration of Conformity Nitrogen N 10DR Nitrogen Generator UM N 10DR Rev 4...

Страница 6: ...0 01 m Outlet 1 4 BSP 1 Pressure Gauges 1 Pthalates None Suspended Liquids None Electrical Requirements 230V 50 60 Hz 8 4 amps Compressor fuse on Control PCB 8 0 amps PLC fan fuse on Control PCB 2 0 a...

Страница 7: ...uble free Nitrogen generation Please review each of the following sections carefully and ensure that the maintenance log at the rear of this manual is updated for future reference Thank you for select...

Страница 8: ...to unpacking Unpacking and Installation ANY DAMAGE SHOULD BE REPORTED IMMEDIATELY TO THE CARRIER AND PEAK SCIENTIFIC OR THE DISTRIBUTOR FROM WHERE THE UNIT WAS PURCHASED Follow the un packing instruct...

Страница 9: ...et to the Compressors Remove Screws 6 7 holding the Bracket to the internal side panels Remove transit bracket from unit Re fit the rear panel in reverse order of removal Important Note The Transit Br...

Страница 10: ...ab i e for the first 10 meters use 6 4 tubing and the final 20 meters use 10 8 Keep the connections and bends to a minimum The imperial equivalents are 6 4 1 4 O D 3 16 I D 10 8 3 8 O D 5 16 I D Elect...

Страница 11: ...imity of Air Conditioning outlets These can sometimes give rise to pockets of air with high relative humidity Operation of the unit within such a pocket could adversely affect its performance Consider...

Страница 12: ...nt panel lock Pressure Gauge Service light Connections Nitrogen N 10DR Nitrogen Generator UM N 10DR Rev 4 Power Cord 230V ONLY 04 1038 EU 04 1029 US 04 1030 UK Rear panel lock 9 way D type connector f...

Страница 13: ...The system will now run through an automatic compressor check procedure Each of the four compressors is tested in turn to ensure that it is working The compressor check takes a minute or so and then...

Страница 14: ...0 days MENU Software Rev V1 55 System Hours Diagnostic To display the current status of each compressor the system and the inlet filter The compressors are numbered 1 to 4 from the right hand side loo...

Страница 15: ...oller Rotate knurled locking nut at base of adjustment knob to release lock Rotate to increase decrease flow Rotate knurled locking nut at base of adjustment knob to lock Unusual Operation If at any t...

Страница 16: ...minutes for unit to fully de pressurise After this time pull out the Quick Release Ring on the Pressure Relief valve situated on the tank Ensure tank is fully depressurised before releasing the ring S...

Страница 17: ...17 Work Done Remarks Name Date Maintenance Log for Serial Number _____________ Maintenance Log Nitrogen N 10DR Nitrogen Generator UM N 10DR Rev 4...

Отзывы: