Start Up Sequence When the unit is switched on the fans should power up and the 
display on the front panel should read –

The  N*10DR  generator  is  designed  specifically  to  minimize  operator  involvement. 

Given that the system is installed as described in earlier sections and is serviced in 

accordance  with  the  following  maintenance  recommendations  then  it  should  simply 

be a matter of turning the generator on. The generator will automatically produce the 

factory default flow and pressure.

Gas Generation When the system is running, the maximum number of compressors 

operating at any time is two. To ensure the longevity of the system, the compressors 

cycle in pairs, with a pair running for a three minutes while the other pair are turned 

off to cool, then the running pair are switched off to cool and the idle pair are switched 

on. The display will read -

Compressor Test

The system will now run through an automatic compressor check procedure. Each of 

the four compressors is tested in turn to ensure that it is  working. The compressor 

check takes a minute or so and then the system will turn on fully and “PLEASE WAIT”

will be displayed while pressure builds. The pressure can be monitored on the front 

panel gauge.

*Note: The outlet should be vented to atmosphere for at least 30 seconds to 

purge the system of impurities.

If only a small volume of gas is being drawn from the system, it will automatically 

change into ESI mode. The display will read -

Peak Scientific

Mode: APCI

Whilst running in this mode, if too much gas is then being drawn from the system, it 

will automatically change into APCI mode to meet the increased demand.

The  user  can  manually  select  ESI or  APCI Compressor  mode  by  pressing  the 

relevant  button  on  the  control  panel.  If  ESI mode  is  selected  while  the  system 

requires two compressors, it will automatically return to APCI mode.

Peak Scientific

Mode: ESI

ESI

APCI

Operation

Nitrogen N*10DR Nitrogen Generator

UM  – N*10DR – Rev 4

Содержание N 10DR Series

Страница 1: ...1 User Manual UM N 10DR Rev 4 Nitrogen N 10DR Nitrogen Generator...

Страница 2: ...duction 7 Unpacking and Installation 8 Removal of Transit Bracket 9 Useful Information 10 Electrical Connection 10 Generator Environment 11 General Dimensions 11 Generator Connections 12 Generator Con...

Страница 3: ...elease G L 16 10 06 2 Notes added for Indoor use only Altitude T G 10 09 07 3 Update USA address S K 25 08 2008 4 Updated solenoid valve part S K 26 01 2009 Document Change History Nitrogen N 10DR Nit...

Страница 4: ...5 Subject to the Company being notified of any defect as is referred to in sub clause 2 hereof within a reasonable time of it becoming apparent and subject always to the terms of sub clause 4 hereof...

Страница 5: ...5 Declaration of Conformity Nitrogen N 10DR Nitrogen Generator UM N 10DR Rev 4...

Страница 6: ...0 01 m Outlet 1 4 BSP 1 Pressure Gauges 1 Pthalates None Suspended Liquids None Electrical Requirements 230V 50 60 Hz 8 4 amps Compressor fuse on Control PCB 8 0 amps PLC fan fuse on Control PCB 2 0 a...

Страница 7: ...uble free Nitrogen generation Please review each of the following sections carefully and ensure that the maintenance log at the rear of this manual is updated for future reference Thank you for select...

Страница 8: ...to unpacking Unpacking and Installation ANY DAMAGE SHOULD BE REPORTED IMMEDIATELY TO THE CARRIER AND PEAK SCIENTIFIC OR THE DISTRIBUTOR FROM WHERE THE UNIT WAS PURCHASED Follow the un packing instruct...

Страница 9: ...et to the Compressors Remove Screws 6 7 holding the Bracket to the internal side panels Remove transit bracket from unit Re fit the rear panel in reverse order of removal Important Note The Transit Br...

Страница 10: ...ab i e for the first 10 meters use 6 4 tubing and the final 20 meters use 10 8 Keep the connections and bends to a minimum The imperial equivalents are 6 4 1 4 O D 3 16 I D 10 8 3 8 O D 5 16 I D Elect...

Страница 11: ...imity of Air Conditioning outlets These can sometimes give rise to pockets of air with high relative humidity Operation of the unit within such a pocket could adversely affect its performance Consider...

Страница 12: ...nt panel lock Pressure Gauge Service light Connections Nitrogen N 10DR Nitrogen Generator UM N 10DR Rev 4 Power Cord 230V ONLY 04 1038 EU 04 1029 US 04 1030 UK Rear panel lock 9 way D type connector f...

Страница 13: ...The system will now run through an automatic compressor check procedure Each of the four compressors is tested in turn to ensure that it is working The compressor check takes a minute or so and then...

Страница 14: ...0 days MENU Software Rev V1 55 System Hours Diagnostic To display the current status of each compressor the system and the inlet filter The compressors are numbered 1 to 4 from the right hand side loo...

Страница 15: ...oller Rotate knurled locking nut at base of adjustment knob to release lock Rotate to increase decrease flow Rotate knurled locking nut at base of adjustment knob to lock Unusual Operation If at any t...

Страница 16: ...minutes for unit to fully de pressurise After this time pull out the Quick Release Ring on the Pressure Relief valve situated on the tank Ensure tank is fully depressurised before releasing the ring S...

Страница 17: ...17 Work Done Remarks Name Date Maintenance Log for Serial Number _____________ Maintenance Log Nitrogen N 10DR Nitrogen Generator UM N 10DR Rev 4...

Отзывы: