14
If there is a problem with the system, the user will be informed by one of the two available indicators -
SERVICE : Audible buzzer and light on the front panel
ALARM : Audible buzzer and light on the front panel.
To stop the audible buzzer press on the Control Panel. Note the light will stay illuminated
as long as the unit has power.
MENU
ENTER
Diagnostic
Mode
Compressor 1: OK
Compressor 2: OK
Compressor 3: OK
Compressor 4: OK
System: OK
Inlet filter: OK
MENU
ENTER
MENU
Operating Hours
Diagnostic
ENTER
Comp 1: 29Hr 24Min
Comp 2: 27Hr 30Min
Comp 3: 29Hr 24Min
Comp 4: 27Hr 30Min
MENU
ENTER
System Hours
Operating Hours
Time since last
service: 21 days
Total system run
time: 30 days
MENU
Software Rev. V1.55
System Hours
Diagnostic: To display the current status of each compressor, the system and the inlet filter.
The compressors are numbered 1 to 4 from the right-hand side looking from the rear and are
identified by holes in the panel they are sitting on.
Operating hours: To display the current operating hours of each compressor / membrane
in hours and minutes.
System hours: To display the time since the last service and the total system
operating time in days.
To display the software revision.
To exit.
MENU
ENTER
To exit.
MENU
ENTER
To exit.
MENU
ENTER
To exit.
Control Panel / Fault Diagnosis
Mode: To display the current operating mode.
ENTER
MENU
Mode
Diagnostic
Peak Scientific
Mode: ESI
ALARM
RESET
Interrogating the Control Panel to find fault -
Memb 1: 29Hr 24Min
Memb 2: 27Hr 30Min
Operation
Nitrogen N*10DR Nitrogen Generator
UM – N*10DR – Rev 4