The Generator is designed for indoor use only.
The unit should be installed adjacent to the Mass Spectrometer it is supplying. If this
is not convenient then the unit can be sited elsewhere, however, consideration should
be made of the lengths of pipe runs as pressure drops can result from extended runs
of pipe. Please see the “Useful Information” section for further details.
Performance of the generator (like all sophisticated equipment) is affected by ambient
conditions. Note should also be taken to the proximity of Air Conditioning outlets.
These can sometimes give rise to “pockets” of air with high relative humidity.
Operation of the unit within such a pocket could adversely affect its performance.
Consideration should also be given to the air flow around the unit. It is recommended
that an air gap of 75mm (3”) should be maintained down both sides, at the rear and
across the top of the unit. Please refer to the drawing below for the general
dimensions of the unit.
Maximum Ambient Conditions:
25°C (dry bulb) 70%RH (Max)
General Dimensions
Unit must always be placed on a level surface. Failure to do so will effect the
performance of the Generator.
Generator
Environment
Nitrogen N*10DR Nitrogen Generator
UM – N*10DR – Rev 4