Page 10
Technical Specification
Infinity XE 6010 & 6020
Environment
6010
6020
Minimum Operating Ambient Temperature
5°C (41°F)
Maximum Operating Ambient Temperature
30°C (86°F)
Maximum Altitude
3000m
Maximum Relative Humidity
80%
*
Maximum safety ambient temperature 30°C
**
When taken out of storage the generator should be allowed to acclimatize at room temperature for a
minimum of 3 hours before operation.
Inlet Conditions
Minimum Air Inlet Pressure
4.1 bar (60 psi)
Maximum Air Inlet Pressure
9.9 bar (145 psi)
Minimum Air Inlet Flow
Dependant on required flow and purity
Minimum Air Quality
ISO 8573-1:2010 Class [1.4.1]
Generator Outlets
Maximum Pressure Drop (Inlet-Outlet)
0.7bar (10psi)
Maximum Outlet Flow (95% Nitrogen)
130 L/min
260 L/min
Outlet Nitrogen Purity
95 - 99.5%
Start-Up Time
30 minutes
Particles
<0.01
µ
m
Pressure Gauges
2
(Optional) Dry Air Outlet Pressure
Up to 135Psi (depending on inlet supply)
(Optional) Dry Air Outlet Flow
Up to 210 L/min (depending on inlet supply)
Electrical Requirements
Voltage
100-240 V +/-10%
Frequency
50/60 Hz
Current
0.5-0.25A
Input Connection
C20 Plug
Power Cord
C19 socket to local connection (13A min 16A Max)
Pollution Degree
2
Over Voltage Category
II
General
Dimensions cm (inches) WxDxH
41 x 95 x 99.5 (16.2 x 37.4 x 39.2)
Generator Weight Kg (lbs)
80 kg (176.4 lb)
82.5 kg (182 lb)
Shipping Weight Kg (lbs)
92.5 kg (204 lb)
95 kg (210 lb)