Page 14

Installation

Generator Environment

The generator is designed for indoor use only. It should be installed adjacent to 

the application(s) it is supplying. If this is not convenient then the unit can be sited 

elsewhere, however, consideration should be made of the lengths of pipe runs as 

pressure drops can result from extended runs of pipe. 
Performance of the generator is affected by ambient conditions. Please note that the 

generator’s function is primarily to remove Oxygen and Moisture from the generated 

Nitrogen and/ or Dry Air. Though, depending on model, some hydrocarbon 

technology may be employed by the generator, in environments with high ambient 

concentrations of THC additional THC removal filtration may be required, or service 

life of integrated THC traps may be significantly reduced. Note should also be 

taken to the proximity of Air Conditioning outlets. These can sometimes give rise 

to “pockets” of air with high relative humidity. Operation of the unit within such a 

pocket could adversely affect its performance. Consideration should also be given 

to the air flow around the unit. 

It is recommended that an air gap of 100mm (3.9”) 

should be maintained down both sides and at the rear of the unit

. Please refer to the 

drawing on the following page for the general dimensions of the unit.
Please ensure Generator is situated in a well ventilated environment and is positioned 

to allow easy disconnection from the mains supply if required.  

 

Please ensure

 generator is positined in such a way that it can be easily disconnected 

from the mains  if necessary.
Minimum Operating Ambient Temperature: 

 5 °C (41 °F)

Maximum Operating Ambient Temperature: 

30 °C (86 °F)

Inlet Air Quality

The Nitrogen Generator should be connected to an ISO 8573-1:2010 (1:4:1) source of 

compressed air. A minimum pressure of 60 Psig is required for efficient operation of 

the Generator. Any doubts as to the suitability of your compressed air supply should 

be referred to the factory for advice.

A MAXIMUM OIL VAPOUR CONTENT OF 0.01 mg per m^3 is PERMISSIBLE

Air Inlet, Nitrogen Outlet and Air outlet connections are at the rear of the generator. 

The Air inlet connection is a 16mm Push-Fit Fitting. Outlet ports for both Air and 

Nitrogen are 12mm Push-Fit Fitting, with optional depressurising valves included in 

the fittings kit for the nitrogen outlet. The Drain Outlet is a 6mm Push-Fit Fitting.

Содержание Infinity XE 60 Series

Страница 1: ...Infinity XE 60 Series User Manual Copyright Peak Scientific Instruments Ltd 2021 Infinity XE 60 Series User Manual Rev 3 RSID 4129 EN 28 09 2021...

Страница 2: ...ication 10 Unpacking 12 Fittings Kit Contents 13 Installation 14 Generator Environment 14 Inlet Air Quality 14 Generator Overview 15 General Dimensions 15 Rear Connections 16 Electrical Connection 17...

Страница 3: ...been written as a reference document where you can skip to the relevant information Users can refer to the contents page to find the relevant information Please review each of the following sections...

Страница 4: ...s specified within relevant product user manual at all times d your air compressor dryer filtration and oil removal systems must be deemed suitable for use by Peak or a Peak Partner and must be change...

Страница 5: ...y respect or to create any additional obligations or liabilities for any party involved 11 This warranty statement supersedes any and all prior warranty agreements between the parties and constitutes...

Страница 6: ...ted conditions are fully understood or met A CAUTION notice denotes a hazard It calls attention to an operating procedure process or similar which if not correctly performed or adhered to could cause...

Страница 7: ...5 EU EN 61010 1 2010 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use Electromagnetic Compatibility Directive 2014 30 EU EN 61326 1 2013 Electrical Equipment for...

Страница 8: ...l Equipment Safety Regulations 2016 SI 2016 1101 as amended BS61010 1 2010 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use The Electromagnetic Compatibility Reg...

Страница 9: ...Guidance Notes PDF produced by the Department for Business Innovation and skills for the UK and here for Europe All PEAK products that are subject to the WEEE directive are compliant with the WEEE mar...

Страница 10: ...Quality ISO 8573 1 2010 Class 1 4 1 Generator Outlets Maximum Pressure Drop Inlet Outlet 0 7bar 10psi Maximum Outlet Flow 95 Nitrogen 130 L min 260 L min Outlet Nitrogen Purity 95 99 5 Start Up Time...

Страница 11: ...uality ISO 8573 1 2010 Class 1 4 1 Generator Outlets Maximum Pressure Drop Inlet Outlet 1 4bar 20psi Maximum Outlet Flow 95 Nitrogen 390 L min 520 L min Outlet Nitrogen Purity 95 99 5 Start Up Time 30...

Страница 12: ...e Peak Partner from where the unit was purchased Follow the unpacking instructions posted on the side of the crate It will require two people to remove the unit from the shipping crate and to manoeuvr...

Страница 13: ...bing x 3m 3 6mm Push Fit Fitting x 1 4 12mm Push Fit Fitting x 2 5 16mm Push Fit Fitting x 1 6 Tee Fitting x 1 7 Shut Off Valve x 1 8 UK Mains Power Cable x 1 9 EU Mains Power Cable x 3 10 JP Mains Po...

Страница 14: ...round the unit It is recommended that an air gap of 100mm 3 9 should be maintained down both sides and at the rear of the unit Please refer to the drawing on the following page for the general dimensi...

Страница 15: ...o so will affect the performance of the generator WARNING Ensure all inlets and outlets are connected to correct sources and applications All Connections should only be carried out by trained personne...

Страница 16: ...Page 16 Rear Connections...

Страница 17: ...ate The mains supply must be of the stated AC voltage and frequency EARTH GROUND E Green Yellow or Green LIVE L Brown or Black Neutral N Blue or White Electrical Panel Access Changing settings within...

Страница 18: ...optional Dry Air output the same principle applies to both N2 and Air outlets Eco Mode When demand from the instrument stops Infinity XE60 generator will stop demanding air from the external compresso...

Страница 19: ...nd for the nitrogen stop the generator will automatically go into eco mode refer to page 18 shutting off the air inlet Once the demand restarts the generator automatically resumes producing Nitrogen I...

Страница 20: ...is shows live system pressures Information Screen The toggle arrows allow the user to view the two following information screens The initial information screen indicates the total number of membranes...

Страница 21: ...line and program ver sions Settings Screen These screens allow the user to toggle between different flow and purity measure ment units These can be set to the users personal preference at any time To...

Страница 22: ...Depending on output flows the generator will shut down membranes in order to save house air The generator constantly monitors the optimal number of membranes vs output flow and when the flow is lowere...

Страница 23: ...with a fault reference code Toggle the warning indicator in the top right corner of the HMI to view warning code Minor Fault codes A list of minor fault codes for the XE60 is shown below REF CODE 0100...

Страница 24: ...e Toggle the warning indicator in the top right corner of the HMI to view warning code Major Fault codes A list of major fault codes for the XE60 is shown below REF CODE 0001 N2 Pressure Transducer Fa...

Страница 25: ...Page 25 Service alarms When a service is due the HMI will show the following Icon on the screen When a service is overdue the HMI will show the following Icon on the screen...

Страница 26: ...Page 26 When a service is overdue the HMI will show the following Icon on the screen...

Страница 27: ...rvice Requirements Service Schedule Purchase Interval Component Visit 12 Months Infinity XE 60 Series Annual Maintenance Kit www peakscientific com ordering 4 years Infinity XE 60 Series 4 year Mainte...

Страница 28: ...HMI screens will have additional images Home Screen The Home Screen will include this additional bar graph to show Dry Air output pres sure Information Screen The information screen will include the D...

Страница 29: ...to install and setup your generator IQ OQ Certified assurance for applications requiring documented qualification Complete plan Swift response by a Peak Service Engineer within 72 hours planned preven...

Страница 30: ...lure to place the Generator in a safely isolated condition when instructed to do so may lead to personal injury or injury to others and even death 1 Switch generator off 2 Disconnect from air supply 3...

Страница 31: ...tch is illuminated Disconnect power cord from the rear of the Generator Open the right hand panel are check that the circuit breaker is turned on switch in the up position Reconnect power cord Contact...

Страница 32: ...ic com For generators that remain unregistered after 1 month from the shipment date the warranty will be considered active from the date of factory dispatch Complete Plan only Go Online or Complete an...

Страница 33: ...cated in over 20 countries across every continent around the world This allows us to provide an industry leading rapid response service to our customers With Peak Protected your laboratory s productiv...

Отзывы: