Page 22
System Icons
System icons will show in the top right of the screen and give indications on running
conditions, faults or service alarms. A ‘tick’ icon in the top right indicates normal run-
ning conditions with no faults or service requirements.
‘Eco’ mode
Depending on output flows the generator will shut down membranes in order to save
house air. The generator constantly monitors the optimal number of membranes vs
output flow and when the flow is lowered it will begin to close of the feed to certain
membranes. When the demand is increased it will react by turning on any ‘shut down’
membranes. When a generator is running at purity with one or more membranes ‘shut