Page 31
Troubleshooting
Problem
Possible Solution
The Generator will not switch on and the
power switch does not illuminate.
• Ensure power cord is plugged into the
Generator and that the power socket is
turned on.
• Check the fuse in the power cord plug.
• Contact your service provider.
The Generator will not switch on but the
power switch is illuminated.
• Disconnect power cord from the rear of
the Generator. Open the right hand panel
are check that the circuit breaker is turned
on (switch in the up position). Reconnect
power cord.
• Contact your service provider.
The output pressure has dropped below its
original set point.
• Check the pressure regulator(s) inside the
unit has / have not been adjusted.
• Check the HMI for a low pressure warning
• If so, check your air supply does not have
a pressure drop over the length of tubing
and has no leaks.
• Contact your service provider.