Wireless Optical Dissolved Oxygen Sensor Replacement Sensor Cap Calibra-
20
013-16057A
SPARKvue Calibration Code Coefficients
1.
I
n SPARKvue, click or touch the Hardware Setup panel
in the data display to open the “Con-
figure Sensors and measurements” window.
2.
Click or touch the “Configure Sensor Icon” to open the Edit Sensor Properties Window.
3.
In the area next to “Cap Coefficient” enter the enter the ten numbers from the sheet with
the Calibration Code
Coefficients for Cap Replacement as a list of 10 numbers separated by spaces.
•
NOTE: Enter the numbers with a space between each number.
•
NOTE: Enter the numbers in the text area exactly as they appear on the sheet with the Calibration Code
Coefficients for Cap Replacement including negative signs, decimal points, and other characters.
4.
Touch or Click “Set Cap Coefficient” to save the coefficients. Click or touch “Done” to close the window.
5.
A small window opens to confirm that the procedure was successful.
Configure
Sensor icon
Enter Cap
Coefficients here