PS-3224
Calibration: See www.pasco.com/calibration
13
013-16057A
3.
In the second step it says “Edit the value:”.Refer to
the table of salinity values (shown previously.) and
enter the correct value in the text area. Click
“Finish”.
4.
Click the Calibration icon in the Tools palette to close
the Calibration panel
SPARKvue Calibration. Remember, see
www/pasco.com/calibration
SPARKvue Calibration Steps
•
Place the Wireless Optical Dissolved Oxygen Sensor into your known standard for 100% dissolved oxygen
saturation.
1.
In SPARKvue, touch or click “Sensor Data” in the Welcome Screen, and connect to the Wireless Optical Dis-
solved Oxygen Sensor.
2.
In the SPARKvue “Sensor
Data Configuration” window,
check that the “DO2
Concentration” measurement
and Temperature
measurement are
checkmarked.
3.
In the “Templates” window
next to the “Sensor Data
Configuration” window, click
one of the Data Display
“Templates” (Graph, Table
and Graph or Digits) to open a data display.