PS-3224
Calibration: See www.pasco.com/calibration
15
013-16057A
6.
In the “Calibrate Sensor: Select Measurement” window, DO
2
Saturation (%) is the default measurement. “1
point (Adjust Slope Only) is the default Calibration Type. Click or touch “Continue” to open the “Calibrate
Sensor: Enter Values” window.
7.
In the “Calibrate Sensor:
Enter Values” window,
“Calibration Point 2” is the
default selection with the
“Standard Value: 100.0000”
and the “Sensor Value:.
panel. NOTE: that the
“Sensor Value” panel will
show the number coming
from the Sensor.
8.
When the Sensor Value
stabilizes, click or touch “Set
Calibration”’
9.
Click or touch “New
Calibration” to see the values
of the Slope and Offset.
10.
Click or touch “OK” to return
to the “Configure Sensors
and measurements” panel.
Live DO
2
Saturation value
coming from the Sensor.