Specification
Value
Dissolved Oxygen (mg/l and% saturation)
Range:
0 - 20 mg/l or 0 - 200% saturation
Accuracy:
±0.1 mg/ or ±1.0% (whichever is greater) after calibration
above 200% ±10%
Resolution:
0.01 mg/l or 0.1% saturation
Temperature (C)
Range:
0 - 50
Accuracy:
0.1 °C
Resolution:
0.1°C
Barometric Pressure (inHg)
Optional Units:
mmHg
Range:
375 - 836 mmHg
Accuracy:
TBD
Resolution:
1 mmHg
Environmental Operating Range
Temperature (C):
0 - 50
Humidity:
0 - 100%
Depth:
10 m
Test Standard:
IP-X8
Optional Atmospheric In Air (% concentration)
Range:
0 - 100%
Accuracy:
Qualitative
Resolution:
1%
Address: PASCO scientific
10101 Foothills Blvd.
Roseville, CA 95747-7100
Web:
www.pasco.com
Phone: +1 916-462-8384 (worldwide)