PS-3224
Replacement Sensor Cap Calibration
19
013-16057A
Capstone Calibration Code Coefficients:
1.
With the sensor “paired” (connected to the computing device) use the data collection software to enter the Cal-
ibration Code Coefficients (ten number) for the replacement Sensor Cap into the sensor.
2. In the Hardware Setup panel, click the “Prop
-
erties” icon ( ) in the lower right corner to
open the Sensor Properties window.
3.
In the “Properties” window, touch or c
lick “Cap
Coefficient
s
”
to select the text panel.
4.
Next, enter the ten numbers from the sheet
with
the Calibration Code
Coefficients for Cap
Replacement as a list of 10 numbers sepa-
rated by spaces.
•
NOTE: Enter the numbers with a space
between each number.
•
NOTE: Enter the numbers exactly as they
appear on the sheet with the Calibration Code
Coefficients for Cap Replacement including
negative signs, decimal points, and other
characters.
5.
When finished, touch or c
lick “
OK
” to close the
Properties window.
SUCCESS
A message appears if the Cap Coefficient has been successfully set.
ERROR MESSAGES
If the message titled “Cap Coefficients failure” appears, it will include a description of the problem.
One error message is “Invalid checksum”. Make sure that the 10 numbers exactly match the ones on the
Calibration Code Coefficients for Cap Replacement sheet. Another message is “Coefficients must be a list of 10
numbers separated by spaces”. Re-enter the numbers with a space between each one.
6.
After successfully entering the Calibration Coefficients, p
erform a one-point or a two-point calibration as
described previously.