PHYSIO-PORT UP
A9951 Revision C
Cleaning, Maintenance, Disposal
8.4 Maintenance
Checks before each use
Before each use, visually check the device and the cables
for signs of mechanical damage.
If you detect damage or impaired functions which may
result in a hazard to the patient, the operator or third
persons, the device must be repaired before it can be
used again.
Technical Safety Inspections
For safety, the device requires regular maintenance. To
ensure functional and operational safety of PHYSIO-
PORT UP, Technical Safety Inspections should be
carried out at least every 2 years.
The checks can be carried out within the framework of a
service agreement; please contact PAR Medizintechnik
Service for details. The device does not require any other
maintenance.
Technical Inspections of the Measuring
System
The non-invasive pressure measurement system of
PHYSIO-PORT UP should be inspected every two
years.
The checks are carried out within the framework of a
service agreement; please contact PAR Medizintechnik
Service for details.
8.5
Disposal of the Product
The products described in this operator
manual must not be disposed as unsorted
municipal waste and must be collected
separately. Please contact an authorized
representative of the manufacturer for
information
concerning
the
decommissioning of your equipment.
The cuffs can be disposed as contaminated
hospital waste.
Calibration Mode
(e.g., to check the pneumatic system for leaks)
Connect a rubber bulb between pressure tubing
and cuff, using a T-adapter.
Roll up cuff tight.
Turn device off and on again after a few seconds,
then wait for time to be displayed.
Push four times: the display indicates "H 4".
Push : the display indicates an internal value
that must be between 25 and 100. If the displayed
value is outside this range, PHYSIO-PORT UP
must be returned for repair.
Push again: the display indicates "0" (the
display now indicates the pressure in mmHg).
Generate a test pressure of 200 mmHg and
measure the pressure decrease after waiting at least
30 seconds. (Pressure decreases between 3 mmHg
and 5 mmHg are typical; a pressure decrease > 6
mmHg indicates a leak and the system needs to be
repaired.)
Push to exit the calibration mode.
Viewing the Firmware Version
Turn on the device and wait for the time to be
displayed.
Push five times: the display indicates "H 5".
Push : the firmware version is indicated, e.g.,
– "30" = firmware version 3.0
Push to end the display.
Caution
These checks shall be carried out by PAR
Medizintechnik or authorized companies.
Caution
These checks shall be carried out by PAR
Medizintechnik or by authorized companies.