16
PHYSIO-PORT UP
A9951 Revision C
Setup
Time and Date
Usually the BP monitors are set to the correct time and
date before delivery. Therefore, the time only needs to
be corrected to change between Standard Time and
Daylight Saving Time.
Setting Time and Date
Briefly switch PHYSIO-PORT UP off and on
again and wait for the time to be displayed.
Push twice: the display indicates "H 2".
Push : The year will be displayed, e.g. "2020".
If the indicated year is correct, confirm it with
or correct it with , then confirm with
The month will be displayed, e.g. "03".
If the indicated month is correct, confirm it with
or correct it with , then confirm with
.
In the same manner, correct day, hour, and minute.
In the end, the time of day will be displayed again.
Selecting the Pressure Unit
Briefly switch PHYSIO-PORT UP off and on again
and wait for the time to be displayed.
Push 8 times: the display indicates "H 8".
Push : the display indicates "mmHg" or "kPa".
Either confirm with or switch to the other
option with , then confirm with .
Measurement Protocols
There is a choice of three different measurement
protocols:
Protocol
Day Phase
(7 a.m. to 10 p.m.)
Night Phase
(10 p.m. to 7 a.m.)
P 1
every 15 minute
every 30 minute
P 2
every 20 minute
every 40 minute
P 3
every 30 minute
every 60 minute
Max. inflation pressure: day phase 250 mmHg
night phase 220 mmHg
Selecting a Measurement Protocol
Briefly switch PHYSIO-PORT UP off and on again
and wait for the time to be displayed.
Push 3 times: the display indicates "H 3".
Push : the display indicates "LLLL" (Selecting a
protocol automatically clears the memory. If you want
to retain the data, switch the BP monitor off.)
Push : the display indicates "P1" (protocol 1).
Either select program 2 or 3 by pushing or
Confirm the selected protocol with .
Enabling or Disabling the Audio Signal
Briefly switch PHYSIO-PORT UP off and on
again and wait for the time to be displayed
Push 7 times: the display indicates "H 7".
Push : the display indicates "0000" when the
audio signal is disabled, and "1111" when it is
enabled.
Either confirm with or switch to the other
option with , then confirm with .