Application 

 

18

 

PHYSIO-PORT UP 

A9951 Revision C

 

 

 

 

 

Place  the  cuff  on  that  arm  of  the  patient  which  is 

used  less  frequently  during  normal  daily  activities: 

on adults about 2 fingers' breadth above the bend of 

the  elbow.  Bending  the  arm  must  not  change  the 

cuff level. 

 

It  is  recommended  to  place  a  hose  made  of  mull 

between arm and cuff. 

Verify that 

 

the  cuff  tubing  points  up  toward  the  shoulder 

(Fig. 4-1) 

 

no  compression  or  restriction  of  connection 

tubing can occur 

 

the side with the  

            label  is  on  the  skin 

(single-use cuffs) 

 

the arrow is located above the brachial or femoral 

artery 

 

the  dashed  white  line  at  the  end  of  the  cuff  is 

located  between  the  two  dashed                        lines 

when  you  close  the  cuff  (if  this  is  not  the  case, 

select another cuff size, see Fig. 4-2), 

 

the  cuff  fits  snugly  around  the  arm  but  does  not 

compress the blood vessels 

 

the  cuff  and  the  PHYSIO-PORT  UP  are  used 

inside  the  ambient  conditions  for  operation  and 

inside the measuring range (s. chapter „Technical 
Specifications“ 

 

Performing a Trial Measurement 

 

Turn  on  PHYSIO-PORT  UP  and  place  it  in  the 

wearable pouch. There is an aperture in the pouch to 

accommodate the cuff connection tube. 

 

Attach  the  pouch  to  the  patient  (shoulder  strap, 

belt).  For  reasons  of  hygiene,  it  is  not  advised  to 

carry the pouch on the bare skin. 

 

Guide the pressure tubing around the patient’s neck 

as a strain relief and connect it to the blood pressure 

cuff  port  of  PHYSIO-PORT  UP  (

2

,  Fig.  2-1).  Do 

not wrap the pressure tubing completely around the 

neck to avoid strangulation of the patient. You must 

hear  the  connector  click  into  place.  Ensure  that  the 

tube  is  not  kinked  or  blocked  during  the 

measurement. 

 

Check that the display indicates the time of day. (If 

the  memory  contains  data  from  a  previous 

procedure, the letter "M" will appear on the display  

 

when  you  turn  on  the  device.  If  you  still  try  to 

initiate  a  measurement,  the  message  "LLLL" 

prompts you to clear the memory. Push         twice 

to delete the data. If you want to retain the data, turn 

off the device instead of pushing           .) 

 

To  avoid  erroneous  measurements,  ensure  that 

the  patient  does  not  move  during  the  trial 

measurement. The patient may stand or sit. 

 

Push           to initiate the first measurement. 

Within  a  few  seconds,  the  device  starts  inflating  the 

cuff. When the inflation pressure has been reached, the 

cuff will gradually be deflated (deflation measurement 

method)  or  the  pressure  will  be  released  quickly 

(inflation  measurement  method).    The  changing  cuff 

pressure is indicated on the display and the letter "M" 

appears  with  each  detected  oscillation.  At  the  end  of 

the measurement, the measured data will be displayed 

 

the systolic readings (S in mmHg or kPa) 

 

the diastolic readings (D in mmHg or kPa) and 

 

the pulse rate (HR/min

-1

). 

If an error code, such as "E 29" (insufficient number of  

oscillations 

detected) 

is 

displayed 

after 

the 

measurement, tighten the cuff a little and push  
again (see also chapter „Error Codes). 

If  the  trial  measurement  has  been  completed 

successfully,  the  device  is  ready  for  automatic 

measurements.

Содержание PHYSIO-PORT UP

Страница 1: ...PHYSIO PORT UP Ambulatory Blood Pressure System Firmware Version 3 0 Operator s Manual A9951 ENG Revision C...

Страница 2: ...anual only applies to PHYSIO PORT UP firmware version 3 0 It does not apply to earlier firmware versions Due to continuing product innovation specifications in this manual are subject to change withou...

Страница 3: ...romagnetic Compatibility EMC 29 Revision History This manual is subject to the GE Healthcare change order service The revision code a letter that follows the document part number changes with every up...

Страница 4: ...ecified in CISPR11 EN 55011 class B The CE marking covers only the accessories listed in the Order Information chapter This manual is an integral part of the equipment It should be available to the eq...

Страница 5: ...General Information 5 PHYSIO PORT UP A9951 Revision C PAR Medizintechnik GmbH Co KG Rigistr 11 12277 Berlin Germany Tel 49 30 235 07 00 The country of manufacture is indicated on the device label...

Страница 6: ...g a novel and faster technology already during inflation of the cuff inflation measurement method The deflation measurement method is the most com mon method used With this technique the cuff is infla...

Страница 7: ...g diastolic pressure at 83 mmHg With both methods a pressure transducer measures the cuff pressure as well as the superimposed pressure pulsations During blood pressure measurements the cuff must be a...

Страница 8: ...erators or third parties The responsible organization should identify analyze evaluate and control these risks Changes to the IT network could introduce new risks that require additional analysis Chan...

Страница 9: ...s they may emit higher levels of electromagnetic radiation Caution Equipment damage risk to persons Before connecting the battery charger to the power line check that the voltage ratings on the namepl...

Страница 10: ...ment method deflation measurement method or inflation measurement method 1 Button push to display the most recent para meter readings The display will show systolic value S unit mmHg or kPa shown on t...

Страница 11: ...water No protection against contact and ingress of objects and protection against dripping water when tilted at 15 Keep dry Temperature limits Humidity limits Air pressure limits USB port connection...

Страница 12: ...e replaced by a service technician The capacity of two fully charged or new batteries is sufficient for up to 48 h of operation or 400 blood pressure measurements The capacity of rechargeable batterie...

Страница 13: ...compartment and close Charging NiMH Batteries If PHYSIO PORT UP is powered by rechargeable batteries 4 of them are shipped with the equipment they should be recharged immediately after use 24 hours Us...

Страница 14: ...N Insert charged batteries OR briefly press To switch OFF Remove the batteries or press for 3 seconds Performance Check When turned on PHYSIO PORT UP runs a self test that includes all symbols and seg...

Страница 15: ...layed Push the display indicates H 1 Push the display indicates LLLL To delete the data push again the display indicates 0000 followed by the time if you do not wish to clear the memory turn off the B...

Страница 16: ...nfirm with or switch to the other option with then confirm with Measurement Protocols There is a choice of three different measurement protocols Protocol Day Phase 7 a m to 10 p m Night Phase 10 p m t...

Страница 17: ...cuff pressure due to connection tubing kinking Too frequent measurements can cause injury to the patient due to blood flow interference The application of the cuff over a wound can cause further injur...

Страница 18: ...of PHYSIO PORT UP 2 Fig 2 1 Do not wrap the pressure tubing completely around the neck to avoid strangulation of the patient You must hear the connector click into place Ensure that the tube is not ki...

Страница 19: ...rable pouch to remove the pressure hose only in emergency situations see warning below that the cleaning may only be carried out by qualified medical personnel and not by the patient Absolute Contrain...

Страница 20: ...consecutive unsuccessful measurements Error codes E02 battery depleted E06 inflation time over and E08 maximum number of pressure measurements taken 200 or 400 do not lead to a second measurement The...

Страница 21: ...on PHYSIO PORT UP and wait for the time to be displayed For more information about data output please refer to the Operator s Manual of PhysioPortWin When you have finished downloading data to Physio...

Страница 22: ...d be identified PHYSIO PORT UP does not measure diastolic pressures below 40 mmHg E 15 Motion artifact during diastole detection E 17 Internal hardware error Please contact your local authorized deale...

Страница 23: ...B 1 1 2 0 or 3 0 7 3 USB Driver You will need administrator privileges for installation 1 Turn on the PC and the monitor Exit ALL programs 2 Insert the storage device USB stick with the USB driver If...

Страница 24: ...If a USB serial port greater than COM4 has been selected one of the ports COM1 through COM4 must be disabled at the PC so that it can be assigned to the USB port 3 Select one of the ports COM1 throug...

Страница 25: ...aton rapid Sporicidin or Cidex After disinfection rinse the cuff thoroughly with tap water and air dry Moreover spray disinfection has proved successful after the usage of the cuff Incidin Foam or equ...

Страница 26: ...Please contact an authorized representative of the manufacturer for information concerning the decommissioning of your equipment The cuffs can be disposed as contaminated hospital waste Calibration Mo...

Страница 27: ...suring Method Oscillometric selectable measurement method deflation measurement method or inflation measurement method Battery Charger protection class II IP20 100 to 240 VAC 50 60 Hz 0 5 A Ambient Co...

Страница 28: ...A867 Connection cable PHYSIO PORT UP to PC USB length approx 1 5 meters S2010 Connection cable PHYSIO PORT UP to PC RS232 length approx 1 2 meters A20931 Battery charger with USB connection cable and...

Страница 29: ...RF emissions to EN 55011 CISPR 11 Group 1 PHYSIO PORT UP uses RF energy only for their internal function Therefore their RF emissions are very low and are not likely to cause any interference in near...

Страница 30: ...cal commercial or hospital environment Surge to EN 61000 4 5 IEC 61000 4 5 0 5 kV differential mode 1 0 kV differential mode 0 5 kV common mode 1 0 kV common mode 2 0 kV common mode not applicable not...

Страница 31: ...and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electro magnetic site surveya should be less than the compliance level in each f...

Страница 32: ...mum Output Power of Transmitter W Separation Distance According to Frequency of Transmitter m 150 kHz to 80 MHz d 1 17 P 80 MHz to 800 MHz d 1 17 P 800 MHz to 2 7 GHz d 2 33 P 0 01 0 12 0 12 0 24 0 1...

Страница 33: ...the device from the wearable pouch Please wear the pouch over your clothes You do not have to clean the device after the long term measurement Sometimes the device internally stops the long term measu...

Страница 34: ...iew 27 Functional Description 8 G General Information 5 I Indicators 11 Inflation measurement 7 Inflation Measurement Method 21 Inflation Measurement Technology 7 Information for patients 20 Intended...

Страница 35: ...51 Revision C T Technical Inspections of the Measuring System 27 Technical Safety Inspections 27 Time set 17 Toggling Between Day and Night Phase 21 Trial Measurement 19 U USB Driver installation 24 W...

Страница 36: ...36 PHYSIO PORT UP A9951 Revision C Index PAR Medizintechnik GmbH Co KG Rigistr 11 12277 Berlin Germany Tel 49 30 2350700 www par berlin com Printed in Germany...

Отзывы: