22
PHYSIO-PORT UP
A9951 Revision C
Cleaning, Maintenance, Disposal
6
Error Codes
E 02
Batteries depleted. Code appears when the
battery capacity is insufficient for new BP
measurements. The device differentiates
between two states: the memory has just been
cleared (i.e., the battery test is performed with
a higher drain to ensure that fresh batteries
will be inserted at the beginning of the
measurement) or measurements have already
been taken.
E 03
Measurement time over. Code is displayed
after measurement duration of 180 seconds.
E 06
Inflation time over. The maximum inflation
time of 130 seconds has elapsed. This
condition indicates a leak in the cuff or tubing,
or a defect at the cuff connector
E 07
This code appears
when the device could not determine a
systolic value although the cuff pressure
was already increased twice
when the current cuff pressure would
exceed the selected maximum pressure.
E 08
400 pressure measurements taken; storage
capacity exhausted.
E 14
Diastolic reading below 40 mmHg. Code
appears when the cuff pressure has dropped to
40 mmHg and no diastolic pressure could be
identified (PHYSIO-PORT UP does not
measure diastolic pressures below 40 mmHg).
E 15
Motion artifact during diastole detection.
E 17
Internal hardware error. Please contact your
local authorized dealer (http://www.par-
berlin.com).
E18
Systolic reading outside measuring range.
E 19
Diastolic reading outside measuring range.
(Codes
E 18
and
E 19
are displayed when the
systolic and diastolic values are outside the
range in which oscillations were detected.)
E 21
Difference between systolic and diastolic
pressure too small (10 mmHg or less).
E 22
Motion artifact during systole detection.
E 26
Systolic reading below measuring range.
E 27
Systolic reading above measuring range.
E 29
Insufficient number of oscillations detected:
For a correct measurement, the system must
detect a minimum of 8 oscillations.
For
deflation measurement method
:
Tighten the cuff so that one finger, but not
two, can be inserted between the patient's arm
and the cuff. At the same time the device
switches to a deflation rate of 4 mmHg/s.
When it detects more than 13 oscillations later
on, the rate changes to 6 mmHg/s.
For
inflation measurement method
:
This error message will not be displayed
because PHYSIO-PORT UP automatically
switches to the deflation measurement method
if the number of detected oscillations is
insufficient.