5.5.5 Sensitivity Settings
■
Overview
You can adjust measurement sensitivity according to the light transmittance and measurement
state of the measured object.
For sensitivity settings, you can set a judgment level and judgment filter. You can cope with
various applications by combining judgment level and judgment filter settings.
Judgment level
Adjusting the judgment level enables objects with high transmittance to be measured stably.
The judgment level is a set value that is used to judge whether the sensor head receiver is in a
light entry state or light interception state according to the intensity of light received by the
receiver.
If the intensity of received light that is registered as the reference waveform is regarded as
100% and the intensity of received light falls below a judgment level of 25% (default value), the
receiver will be judged to be in a light interception state and the edge position will be calculated.
For measured objects with high transmittance, you must increase the judgment level because
light is received even in sections where light is intercepted.
When the measured object is opaque
When the measured object is translucent
5.5 Settings on the Base Screen (Using Shortcut Keys)
5-40
WUME-HGT-6
Содержание HG-T Series
Страница 2: ... MEMO 2 WUME HGT 6 ...
Страница 28: ... MEMO 2 8 WUME HGT 6 ...
Страница 96: ... MEMO 5 50 WUME HGT 6 ...
Страница 99: ... Note 2 Displayed only during calculation 6 1 Flow of Measurement Data WUME HGT 6 6 3 ...
Страница 128: ... MEMO 6 32 WUME HGT 6 ...
Страница 158: ... No timer OFF 7 2 Advanced Settings 7 30 WUME HGT 6 ...
Страница 177: ...Example Positioning a glass plate Setting method 7 2 Advanced Settings WUME HGT 6 7 49 ...
Страница 218: ...8 2 3 Side View Attachment HG TSV10 Units mm 8 2 Dimension Drawings 8 14 WUME HGT 6 ...
Страница 221: ...10 Troubleshooting 10 1 Troubleshooting 10 2 10 2 Error Messages 10 5 WUME HGT 6 10 1 ...
Страница 228: ...Menu Structure General Function Display Menu Structure General Function Display App 2 WUME HGT 6 ...
Страница 229: ...Menu Structure General Function Display WUME HGT 6 App 3 ...
Страница 230: ...Menu Structure Extended Function Display Menu Structure Extended Function Display App 4 WUME HGT 6 ...
Страница 231: ...Menu Structure Extended Function Display WUME HGT 6 App 5 ...
Страница 232: ... MEMO App 6 WUME HGT 6 ...
Страница 235: ... MEMO WUME HGT 6 ...
Страница 236: ...Panasonic Industrial Devices SUNX Co Ltd 2020 PRINTED IN JAPAN WUME HGT 6 ...