910-257C
March 2006
Page 4-30
MG2 Series
gain, the displayed waveform will be not be ADC’d and
will be shown at an amplitude proportional to the gain
setting.
To return to default ADC’D gain:
Step 1: While in Waveform Adjust Mode, use the [
] or [
] key to
highlight Gain.
Step 2: Press [2
nd
F],[
] (Gain). The gain will be changed to its
default ADC’d condition.
4.13.2 Extended Blank
Extended Blank allows the operator to extend a blank from the sound
entry point to the maximum of the displayed range. When the
extended blank is set, the gage measures to the first valid echo after
the end of the blank.
While the Live Waveform is active, pressing [2
nd
F],[
] (Gain)
places the gage in Waveform Adjust Mode. The user can use the [
]
or [
] key to select Extended Blank and the [
,
] keys to
increase or decrease the blank position. When the blank is set, the
gage measures to the first echo after the end of the blank that is higher
in amplitude than the first gradicule’s 20% screen height.
Содержание MG2
Страница 1: ...USER S MANUAL MG2 MG2 XT MG2 DL Ultrasonic Thickness Gages Part No 910 257C...
Страница 14: ...910 257C March 2006 Page 1 6 MG2 Series...
Страница 31: ...March 2006 910 257C MG2 Series Page 3 13 within the thickness range of 0 100 0 500 2 5 mm 12 5 mm...
Страница 32: ...910 257C March 2006 Page 3 14 MG2 Series...
Страница 80: ...910 257C March 2006 Page 4 48 MG2 Series...
Страница 107: ...March 2006 910 257C MG2 Series Page 5 27...
Страница 108: ...910 257C March 2006 Page 5 28 MG2 Series...
Страница 114: ...910 257C March 2006 Page 6 6 MG2 Series...
Страница 116: ...910 257C March 2006 Page 7 2 MG2 Series...
Страница 138: ...910 257C March 2006 Page I 4 MG2 Series...