March 2006
910-257C
MG2 Series
Page 3-5
Figure 3-2: Unknown Sound Material Do Screen
Step 1:
Update the Transducer Zero Compensation by wiping the
transducer face clean of all couplant and pressing [2
nd
F],
[Cal Zero](Do Zero).
Step 2:
Couple the transducer to the block.
Step 3:
Press the [CAL VEL] key.
Step 4:
When the thickness reading is stable, press the [ENTER]
key.
Step 5:
Uncouple the transducer and use the [
,
,
,
and ]
slewing keys to enter the thickness of the standard.
Step 6:
Press the [MEAS] key to complete the calibration and return
to Measure mode.
Содержание MG2
Страница 1: ...USER S MANUAL MG2 MG2 XT MG2 DL Ultrasonic Thickness Gages Part No 910 257C...
Страница 14: ...910 257C March 2006 Page 1 6 MG2 Series...
Страница 31: ...March 2006 910 257C MG2 Series Page 3 13 within the thickness range of 0 100 0 500 2 5 mm 12 5 mm...
Страница 32: ...910 257C March 2006 Page 3 14 MG2 Series...
Страница 80: ...910 257C March 2006 Page 4 48 MG2 Series...
Страница 107: ...March 2006 910 257C MG2 Series Page 5 27...
Страница 108: ...910 257C March 2006 Page 5 28 MG2 Series...
Страница 114: ...910 257C March 2006 Page 6 6 MG2 Series...
Страница 116: ...910 257C March 2006 Page 7 2 MG2 Series...
Страница 138: ...910 257C March 2006 Page I 4 MG2 Series...