MG2 Series
Page 3-3
[2
nd
F], [Cal Zero](Do Zero) should be pressed to compensate for any
temperature changes in the transducer.
3.3 Velocity and Zero
Calibration
The Material Velocity and Zero Calibration procedures may be
combined using a thick and a thin calibration block of the same
material.
Step 1:
Update the Transducer Zero Compensation by wiping the
transducer face and pressing [2
nd
F], [Cal Zero](Do Zero).
Step 2:
Couple the transducer to the thick calibration block.
Step 3:
Press [CAL VEL].
Step 4:
When the thickness reading is stable, press the [ENTER]
key.
Step 5:
Uncouple the transducer from the block and use the [
,
,
,
and
] slewing keys to enter the thickness of the
thick block.
Step 6:
Couple the transducer to the thin block and press the [CAL
ZERO] key.
Step 7:
When the reading is stable, press the [ENTER] key.
Step 8:
Uncouple the transducer from the block and use the [
,
,
,
and
] slewing keys to enter the thickness of the thin
block.
Step 9:
Press the [MEAS] key to complete the calibration and go to
the Measure mode.
Содержание MG2
Страница 1: ...USER S MANUAL MG2 MG2 XT MG2 DL Ultrasonic Thickness Gages Part No 910 257C...
Страница 14: ...910 257C March 2006 Page 1 6 MG2 Series...
Страница 31: ...March 2006 910 257C MG2 Series Page 3 13 within the thickness range of 0 100 0 500 2 5 mm 12 5 mm...
Страница 32: ...910 257C March 2006 Page 3 14 MG2 Series...
Страница 80: ...910 257C March 2006 Page 4 48 MG2 Series...
Страница 107: ...March 2006 910 257C MG2 Series Page 5 27...
Страница 108: ...910 257C March 2006 Page 5 28 MG2 Series...
Страница 114: ...910 257C March 2006 Page 6 6 MG2 Series...
Страница 116: ...910 257C March 2006 Page 7 2 MG2 Series...
Страница 138: ...910 257C March 2006 Page I 4 MG2 Series...