910-257C
March 2006
Page 4-26
MG2 Series
Waveform Display
To activate the optional Waveform, use the following procedure:
Step 1: Press [2
nd
F], [
] (Setup) to display the Setup tabs.
Step 2: Use the [
,
] keys to select the Waveform tab.
Step 3: Press [
] or [
] to scroll to Enable.
Step 4: Press [
,
] to choose On or Off.
Step 5: Press [
] or [ENTER] to scroll to Zoom.
Step 6: Press [
,
] to choose On or Off.
Step 7: Press [
] or [ENTER] to scroll to Save Key (MG2-DL
only).
Step 8: Press [
,
] to choose THK ONLY or THK + WF.
Step 9: Press [
] or [ENTER] to scroll to Rectification.
Step 10: Press [
,
] to choose among Full, Half Neg, Half Pos,
and RF.