March 2006
910-257C
MG2 Series
Page 4-27
Step 11: Press [
] or [ENTER] to scroll to Waveform Fill.
Step 12: Press [
,
] to choose between Off and On.
Step 13: Press [MEAS/RESET].
Note: When the Waveform has been enabled and the Zoom
feature is turned off, the user can adjust the range and
delay of the waveform by using the [
] or [
] key to
change the range to the next higher or lower range set
point or the [
,
] keys to increase or decrease the
waveform delay.
4.13 Waveform Adjust Mode (MG2-XT
and MG2-DL Only)
The Waveform Adjust Mode feature allows the user to make Setup
adjustments that are not available in the Standard Mode.
Additional measurement capabilities include:
• Manual Gain adjustment in 1 dB increments
• Extended Blank
• Echo Blank (Manual Echo-to-Echo only)
• Range
• Delay
Note: These features can be activated only while the optional