910-257C
March 2006
Page 4-28
MG2 Series
Waveform is enabled. These additional features are not
available when the Waveform is turned off.
Note: These adjustments can affect the measurement capabil-
ity of the instrument and should be undertaken only by a
person with extensive knowledge of ultrasonic theory.
To activate Waveform Adjust Mode while Waveform is enabled:
Step 1: Press [2
nd
F], [
](Gain).
Step 2: Press [
] or [
] to select the waveform setup parameter.
Step 3: Press [
,
] to adjust the parameter.
Step 4: Press [2
nd
F] (Gain) while viewing the Gain parameter to set
the Gain back to its default value.
Step 5: Press [MEAS] to exit the Waveform Adjust Mode with the
parameters at their adjusted values.
4.13.1 Manual Gain Adjust (MG2-XT and
MG2-DL Only)
When the waveform is not activated or if the waveform option was not
purchased, pressing [2
nd
F],[
] (Gain) will move through Gain Hi,
Low, and Standard.
While the live waveform is active, pressing [2
nd
F],[
] (Gain) will
place the gage in Waveform Adjust Mode. The user can use the [
]
or [
] key to select Gain and the [
,
] keys to increase or
decrease the gain. The gain can be adjusted in 1 dB increments from
21 dB to 77 dB. Note that echoes that are higher in amplitude than the
Содержание MG2
Страница 1: ...USER S MANUAL MG2 MG2 XT MG2 DL Ultrasonic Thickness Gages Part No 910 257C...
Страница 14: ...910 257C March 2006 Page 1 6 MG2 Series...
Страница 31: ...March 2006 910 257C MG2 Series Page 3 13 within the thickness range of 0 100 0 500 2 5 mm 12 5 mm...
Страница 32: ...910 257C March 2006 Page 3 14 MG2 Series...
Страница 80: ...910 257C March 2006 Page 4 48 MG2 Series...
Страница 107: ...March 2006 910 257C MG2 Series Page 5 27...
Страница 108: ...910 257C March 2006 Page 5 28 MG2 Series...
Страница 114: ...910 257C March 2006 Page 6 6 MG2 Series...
Страница 116: ...910 257C March 2006 Page 7 2 MG2 Series...
Страница 138: ...910 257C March 2006 Page I 4 MG2 Series...