4.Advanced User Guidebook
81
waveform appears on the screen.
2. Select
Waveform Math
in the bottom menu.
3. In the right menu, select
Factor1
as
CH1
.
4. Select
Sign
as
+
in the right menu.
5. In the right menu, select
Factor2
as
CH2
.
6. Select
Vertical (div)
in the right menu, click
or
to adjust the vertical
position of Math waveform.
7. Select
Vertical (V/div)
in the right menu, click
or
to adjust the
vertical division of Math waveform.
User defined function
1. Click the
Math
softkey on panel to switch to the ON state and display the
math menu on the bottom.
2. Select
User Function
in the bottom menu, select
Edit fun
in the right
menu, an expression input keyboard pops up.
Channel
Confirm
Clear
Operators
Integral
Differential Square root
Expression
3. Create an expression. When done, choose
in the keyboard to confirm.
The division of Math waveform is displayed at the left bottom of screen.
Digital Filter
Digital filter provides 4 types of filters (low pass, high pass, band pass and
band reject). The specified frequencies can be filtered by setting the cut-off
frequency. Digital filter can only apply to CH1 or CH2.
1. Click the
Math
softkey on panel to switch to the ON state and display the
math menu on the bottom.
2. Select
DIR
in the bottom menu.
3. In the right menu, select
channel
as
CH1
or
CH2
.
Содержание TAO3000 Dual-Channel Series
Страница 1: ...TAO3000 Dual Channel Series Tablet Oscilloscopes User Manual www owon com ...
Страница 93: ...4 Advanced User Guidebook 87 Figure 4 24 Hamming window Figure 4 25 Rectangle window ...
Страница 94: ...4 Advanced User Guidebook 88 Figure 4 26 Blackman window Figure 4 27 Hanning window ...
Страница 95: ...4 Advanced User Guidebook 89 Figure 4 28 Kaiser window Figure 4 29 Bartlett window ...