Verify Proper Operation of the Equipment, cont.
Connect the power meter and light source together with the mandrel-wrapped patch cable as shown in Figure 4 below.
Compare the reading on the WaveTester to Table 2 for acceptable power levels. The patch cable should be replaced if it
is below the acceptable power level.
Once proper operation has been verified, continue on to the next step.
Set Reference
Setting a reference is commonly referred to as “zeroing”.
With the power meter and light source still connected as shown in Figure 4, press and hold
on the WaveTester to set
the reference for 850nm.
The WaveTester display should change to show readings in ‘dB’, and the power reading should be approximately 0.00
dB. This means that the meter and light source have been successfully “zeroed” at 850nm.
If testing at 1300nm is also required, connect a mandrel-wrapped patch cable to the 1300nm port, then
UNITS
ZERO