6.6 Installation Procedure
Install the meter body in the following manner:
6.6.1 Wafer Type
① Install adjust rings on the meter periphery at both ends. ASME 150 and JPI 150 meters 25 millimeters in
nominal diameter do not require adjust rings.
② Fitting flange gaskets at both sides, sandwich the meter body between flanges.
CAUTION: Exercise care not to allow flange gaskets to extrude into the interior of the meter (pipe), or
meter accuracy will suffer.
③ Pass the connecting bolts through their
openings and tighten up the bolts. Individual
bolts are now in contact with the adjust rings
and the inner wall of flange bolt holes, forcing
the meter body to be concentric to the pipeline.
IMPORTANT Unless adjust rings are installed in
place, flowmeter-to-pipe misalignment will
result in loss of meter accuracy. Be sure, for this
reason, to fit the adjust rings in place before
bolting the flanges.
Nominal Dia.
10mm to 25mm
Nominal Dia.
40mm to 150mm
Nominal Dia.
10mm to 25mm
Nominal dia.
40mm to 150mm