13
D-546-9-E
(3) Space Saving Arrangement
If required straight pipe space is not obtainable upstream of the vortex flowmeter due to existing restrictions
on installation location, the OVAL Honey Vane S combined with a short pipe may be used to get around the
space problem. EX Delta SS with a built-in Honey Vane S requiring no upstream straight pipe is accurate to ±
2% RD in liquid applications. Accuracy varies with the length of short pipe; consult factory for application
assistance.
◦Honey Vane Outline Dimensions
NOTE: The dimensions of honey vane built-in EX Delta (EX Delta SS) remain the same as those of standard
EX Delta, flanged type (see page 49).
5.2 Pipes to be Used
Nominal thickness Sch. 40 pipes should be used for upstream and downstream pipes of this meter.
5.3 Location of Pressure Gauge and Thermometer Taps
➡
➡
Taps for the pressure gauge and/or thermometer, if
desired, should be located as illustrated in the figure
at right.
Honey Vane“L”
Gage Tap
*: Dimensions at JIS 10K.
Dimensions in millimeters
Honey Vane“S”
Flow
Direction
D=Nominal dia.
Meter Body
Thermo-
meter Tap
Содержание EX DELTA, EX DELTA-DIA