8E38=9, 8E39=9, 8E41=9
Display of the MyoSelect (757T13)
Adjustment regulator A is used to adjust the proportional speed
of the “OPENING / CLOSING” function and the speed of grip
force build-up.
Notice:
Always check the adjustments on the prosthesis and
adapt them to the amputee.
9.5 Control mode 5:
Digital Control
Control with 2 electrodes,
1 electrode and 1 switch or
1 switch
The MyoHand VariPlus Speed closes at constant speed, and the grip force increases propor-
tionally to the duration of gripping. The OPENING / CLOSING speed and the speed of grip force
build-up can be adjusted with adjustment regulator B.
9.5.1 Control with 2 electrodes
Opening:
Digital, signal to the OPENING electrode opens the MyoHand VariPlus Speed at
constant speed.
Closing
: Digital, signal to the CLOSING electrode closes the MyoHand VariPlus Speed at
constant speed.
Control
mode 5
OPEN
CLOSE
Indication
Digital
Control
Sustained electrode signal
Sustained electrode signal
For amputees
with two weak
electrode sig-
nals
Grip force
:
Duration of the signal
Adjustment regulator B
Adjustment regulator B
Display of the MyoSelect (757T13)
Adjustment regulator B is used to adjust the constant speed
of the "OPENING / CLOSING" function and the speed of grip
force build-up.
Notice:
Always check the adjustments on the prosthesis and
adapt them to the amputee.