Configuring the OTT netDL/OTT DuoSens for the OTT Pluvio
2
S
with SDI-12 interface
䡵
Create a OTT netDL/OTT DuoSens channel with “SDI-12 Master” or
“OTT SDI RS485” function block (“Serial sensors” tab).
䡵
Make the following settings:
䊳
Terminal block
OTT netDL “SDI-12 Master”: C 3-4 (specified)
OTT netDL “OTT SDI RS485”: C 1-2 (specified)
OTT DuoSens “SDI-12 Master”: A 3-4 (specified)
OTT DuoSens “OTT SDI RS485”: A 1-2 (specified)
Terminal block used (screw terminal strip) of the
OTT netDL/OTT DuoSens
䊳
Slave address
SDI-12 bus address. Each slave address may be
assigned on an SDI-12 bus line only once. (Check/
set: refer to OTT netDL/OTT DuoSens Operating
Instructions, Chapter SDI-12 Transparent Mode.)
Typical setting: “0” (only one OTT Pluvio
2
S is con -
nected to the terminal block, no bus operation).
䊳
Value no.
identifies which value (the xth of n status
information) of the OTT Pluvio
2
S is recorded in this
channel. Typical setting: “1” (first one of nine (“M!”
Measuring Mode) or three (“M1!” Measuring Mode)
measured values)
䊳
Measuring mode
“M!” or “M1!” (for allocating the measured values,
refer Chapter 6, “SDI-12 commands and responses”)
䊳
allocating the additional measured values of the
OTT Pluvio
2
S to virtual terminals (for allocating the
measured values, refer to Chapter 6, “SDI-12 com-
mands and responses”).
䊳
Concurrent mode
only for OTT netDL: refer to the operating program
online help.
䊳
Instantaneous value
only for OTT netDL: refer to the operating program
online help.
䡵
In the respective Channel function blocks, set the required units and number of
digits after the decimal place
Notes:
䊳
For recording all the 12 measured status information for an
OTT Pluvio
2
S, 12 channels in the OTT netDL/OTT DuoSens are thus necessary.
The first channel contains the function block “SDI-12 Master” or “OTT SDI RS485”
as the input signal. The other channels each contain a function block “Virtual
Sensor” (V02 to V09) as the input signal, (V03 in case of M1!). Of course, just
individual channels can also be recorded. In this case, there are fewer entries
required in the “Value no./Virtual Terminal ID” field.
䊳
Further information on the SDI-12 commands and responses used can be found
in Chapter 6, “SDI-12 commands and responses”.
䊳
The OTT Pluvio
2
S makes the measurement results available for retrieval
immediately after the
aM!
and
aM1!
SDI-12 commands.
Fig. A3: Setting the operating parameters
of the OTT netDL/OTT DuoSens SDI-12
Master function block.
The function block OTT SDI RS485
is set in the same way.
(Example shown: OTT DuoSens).
Value no./
Virtual Terminal ID
48