Configuring OTT netDL/OTT DuoSens for OTT Pluvio
2
S with pulse
output
䡵
Create two OTT netDL/OTT DuoSens channels with function block “Pulse input”
(“Digital sensors” tab). (When you only want to record the amount of precipita-
tion, one function block is sufficient.)
䡵
Make the following settings:
䊳
Terminal block OTT netDL: D 1-2, D 3-4, E 1-2 or E 3-4
OTT DuoSens: B 1-2 or B 3-4
䊳
Pulse factor
– amount of precipitation: 0.05 · 0.1 · 0.2 · 0.5 · 1.0
(one pulse corresponds to
0.05 mm, 0.1 mm, 0.2 mm,
0.5 mm or 1.0 mm of precipitation.)
– status information: 1
䊳
only for OTT netDL: time in milliseconds during which the
pulse input is disabled after recording a pulse. This prevents
the OTT netDL from recording unwanted pulses caused by
bouncing switch contacts. Set 8 ms for the OTT Pluvio
2
S.
Please note:
䊳
Always position a pulse input at the beginning (top) of the function tree.
䊳
After changing the pulse factor, a data reset (reset the OTT netDL/OTT DuoSens)
is necessary!
䡵
In the “Channel” function block, set the required unit and number of digits after
the decimal place.
Fig. A6: Setting the operating parameters
of the OTT netDL/OTT DuoSens
Pulse input function block
(example shown: OTT DuoSens).
Debounce
delay [ms]
Fig. A7: Example configuration of an
OTT DuoSens with 2 values recorded.
Additional configuration examples
can be found on the flash drive
OTT Pluvio
2
S software.
50