CAPELLA Cochlear Emissions Analyzer
first change ear, perform a Probe fit and then run the Auto
sequence.
4.
After creating a new Auto sequence, click on
Run
to execute it,
or
Close
to exit from the dialog box.
5.
Any existing macro can be deleted or edited from the Auto
sequence box.
4.12
Compare
Use this function to compare measurements made on the same cli-
ent's ear over time:
1.
Select
Compare
from the
File
menu.
2.
Choose
TEOAE
,
DP-Gram
or
DP-I/O
from the
Mode
menu as
required.
3.
Click on
Open session
in the
File
menu to find the data to be
compared.
4.
Click on the first session to be compared, then on
OK
, and
then select another measurement if required by means of the
left
or
right
arrows in the
Session
sub-window.
5.
Repeat step 4 up to five times. Each successive measurement
result will be assigned a session number and superimposed on
the others on the graph. To display the corresponding meas-
urement parameters and/or statistics, select
Param.
and/or