CAPELLA Cochlear Emissions Analyzer
User Manual
4.11
Auto sequence
To permit automatic testing,
Capella
has a built-in macro function.
1.
First complete
Probe fit
for the ear to be tested, and then
select
Auto sequence
from the
Setup
menu or click on the
Auto sequence
icon.
2.
Select an existing Auto sequence from the box, and click on
Run
to start measurements in the predefined sequence and
with the predefined parameters. The measurement can be
interrupted at any time by clicking on
Pause
, or the Auto
sequence can be terminated prematurely by clicking on the
Stop
button in the
Auto Sequence
run box.
3.
To create a new macro, type a name into the
New Auto
sequence
box and then click on
New
. Select the measure-
ment modes you wish to test with, type the number in the
sequence under
Mode order
, choose the
Number of meas-
urements
you wish to make for each mode, and then set up
each measurement mode by means of
Parameters
. If
required, choose the
Break
condition for Probe fit, Echo-
Screen and Tympanometry. Remember that each Auto
sequence is for one ear at a time, i.e. to test both ears you must