GN Otometrics A/S
33
CAPELLA Cochlear Emissions Analyzer
User Manual
3.
Change the stimulus level by clicking the arrow keys in the
F1
Level
box or pressing the up and down arrow keys on the
keyboard.
4.
Indicate the desired frequencies by clicking in the appropriate
boxes under the DP-Gram.
5.
Click on
Start
to begin measuring. When the stop criteria for
one frequency have been met, measurement will automati-
cally resume on the next preselected frequency.
6.
Click on
Stop
to end the measurement if you have disabled
Stop criteria
. To resume measurement on the next prese-
lected frequency, click on
Start
.
7.
If you wish to halt the measurement momentarily, click on
Pause
. Click on
Cont.
to resume the measurement, at the
same frequency.
8.
Measurement status can be monitored by observing the
Sweeps
and
Statistics
boxes on the lower right part of the
screen. The noise floor during the measurement is also indi-
cated by the blue vertical bar to the right of the DP-Gram.