34
GN Otometrics A/S
User Manual
CAPELLA Cochlear Emissions Analyzer
9.
Use the right and left arrow keys in the
Frequency
box on
screen or on the keyboard to view the amplitude spectra for
different frequencies.
Data for the selected frequency appear in the
Sweeps
and
Statistics
boxes, and in the boxes labeled
DP1
and
DP2
.
10.
The DP1 and DP2 and noise values can be viewed in a tabular
form by choosing
DP Tabular
on the
View
menu. Informa-
tion regarding the f1:f2 ratio, primary and DP frequencies, pri-
mary levels and DP phases can be viewed by choosing
DP
Info
on the
View
menu.
11.
The default pattern showing normative data in the DP-Gram
display is for adults, measured at the University Hospital of
Copenhagen. To show the corresponding pattern for babies
instead, select
Norm. data
in the
Setup
menu and click on
baby
. Amongst the other normative data patterns which may
be selected from this dialog box is the
Gorga
norm. Confirm
the selected norm by clicking on
OK
. The normally green DP-
dot turns yellow if lower than the noise floor or outside the
dynamic range of the instrument.
12.
Up to 6 measurements can be made for each ear. Click on the
arrow keys in the
Meas. No.
box on the lower left to make a
new measurement or to view other measurements made on
the same ear during the session.
13.
Repeat the procedure beginning with Probe fit for measuring
the other ear.