A.
Connector for power cable
Warning
•
To avoid the risk of electric shock, this equipment must only be connected to a mains supply with pro-
tective ground.
2.2.3.4
Disconnecting the Mains power cable
The device and any device to be connected which has its own power supply should be turned off before any connections
are established.
To disconnect the device from the mains supply, pull the mains plug out of the wall mains outlet. Do
not position the unit so that it is difficult to pull the mains plug out of the wall mains.
2.2.3.5
Filling and priming procedure
The ICS AirCal uses water to cool the air. It is required that the water reservoir be filled prior to use of the device.
Warning
•
To prevent electrical shock, the ICS AirCal irrigator must be turned off and disconnected from the electrical
wall outlet when filling or refilling the water reservoir.
Caution
•
Use this procedure if: 1) the ICS AirCal was just received from shipping dry. It needs to be filled and primed,
2) If the water level in the display has gone to zero because user never refilled it when orange bar was present. 3) If the
ICS AirCal was tilted at more than 20° with the water pump running. If none of these criteria were met and the water
level display is showing an orange bar and needing refill, please follow the instructions in
Caution
•
Only
use distilled water. Using common water from the tap will result in mineral deposits and will harm
the ICS AirCal.
Filling
1. Turn off the ICS AirCal irrigator and unplug from the electrical wall outlet.
Otometrics - ICS AirCal
9
2 Getting started