Essential performance
ICS AirCal has no essential performance.
9.1
Accessories and Cables
Item
Part number
Specula (Qty 25)
8-62-44100
Lens (Qty 1)
8-35-33300
Syringe Kit (includes funnel)
8-62-46300
Funnel
1-12-66100
Cradle for Delivery Head
8-62-46200
Manuals CD
8-49-91500
7-pin mini-DIN cable (Chartr 200, Hortmann Vestlab)
8-71-90400
9-pin-subminiature to 7-pin mini-DIN (Chartr XP)
8-71-90300
Foot switch
8-35-39700
Power cord, IEC power hospital grade
7630024
Power cord, standard, EU
8-71-240
Power cord, UK
8-71-80200
10
Notes on EMC (Electromagnetic Compatibility)
•
ICS AirCal is part of a medical electrical system and is thus subject to special safety precautions. For this reason, the
installation and operating instructions provided in this document should be followed closely.
•
Portable and mobile high-frequency communication devices, such as mobile phones, may interfere with the func-
tioning of ICS AirCal.
IEC 60601-1-2:2014 and EN 60601-1-2:2015
Guidance and manufacturer's declaration - electromagnetic emissions for all equipment and systems
ICS AirCal is intended for use in the electromagnetic environment specified below. The user of ICS AirCal should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR11
Group 1
ICS AirCal uses RF energy only for its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic equipment.
Otometrics - ICS AirCal
27
10 Notes on EMC (Electromagnetic Compatibility)