Standards
Safety
Class I, Type BF, IPX0.
Complies with:
IEC 60601-1:2005+A1:2012
IEC 60601-1-6:2010+AMD1:2013
ANSI/AAMI ES60601-1:2005 + A1:2012
CAN/CSA-C22.2 NO. 60601-1:14
CAN/CSA-C22.2 No. 60601-1-6:11(R16)
EMC
IEC 60601-1-2
Performance
Stimulus temperature range
12° to 50°C
Cool stimulus
12° to 37°C (Recommended setting: 24°C)
Warm stimulus
37° to 50°C (Recommended setting: 50°C)
Temperature readout
Digital
Time to temperature
< 60 seconds
Temperature accuracy
+/- 0.4°C
Time range
+/- 0.4°C 1 to 99 seconds (Recommended setting: 60 seconds)
Flow rate range
4 - 10 liters per minute (Recommended setting: 8 liters per minute)
Flow rate accuracy
+/- 0.4 liters per minute
Dimensions
Width
34.3 cm (13.5 inches)
Depth
30.5 cm (12 inches)
Height
10.2 cm (4 inches)
Weight
With empty reservoir
7.9 kg (17.5 lbs)
With full reservoir
8.2 kg (18 lbs)
Power supply
Input voltage
100-240 VAC 50/60 Hz
Power consumption
<192 VA
Fuse
T 1.6H 250V
Applied Part
The applied part of the ICS AirCal is the irrigator head.
26
Otometrics - ICS AirCal
9 Technical Specifications