A.
Funnel
6. Pour water into the funnel to fill the reservoir.
7. Monitor the water level closely. Fill to the bottom of the lip. Do NOT overfill the reservoir.
8. Remove the funnel.
9. Wipe up any excess water that might have spilled into the small indentions where the internal cover is placed.
10. Replace the internal cover.
Priming
1. Plug in the irrigator into the electrical wall outlet and turn on.
2. Turn off the airflow.
3. Locate the priming switch on the back panel
.
4. Press the priming switch to turn on and hold 5 seconds.
5. Release the priming switch to turn off and wait 5 seconds.
6. Repeat steps
and
for a total of 10 times on and 10 times off.
7. Press the cool button
. Decrease the temperature to 12°C.
8. If the temperature reaches the set temperature, then the irrigator is primed. Continue to step
12
Otometrics - ICS AirCal
2 Getting started