default settings have been stored. Remember, the default startup temperature of the ICS AirCal should match the tem-
perature of the first caloric protocol in your list (e.g. if you start with Left Cool - default the AirCal to start with Cool).
3.8
Water level display
The water level of the water in the reservoir is displayed. The
symbol indicates that the water reservoir is full and a
blue bar will display. The
symbol indicates that the water reservoir is half full a blue bar will display. Between half full
and empty the bar will change from blue to orange this indicates that the water reservoir needs to be refilled. And
symbol indicates that the water reservoir is empty.
3.8.1
Refill Procedure:
Caution
•
Do NOT use this refill procedure if: 1) the ICS AirCal was just received from shipping dry. It needs to be filled
and primed, 2) If the water level in the display has gone to zero because user never refilled it when orange bar was
present. 3) If the ICS AirCal was tilted at more than 20° with the water pump running. If one of these criteria were met
please refer to
and follow the
.
Caution
•
Only
use distilled water. Using common water from the tap will result in mineral deposits and will harm
the ICS AirCal.
1. Turn off ICS AirCal and disconnect from the electrical wall outlet.
2. Wipe off any dust on the top of the ICS AirCal prior to opening the case to refill the reservoir.
3. Face the ICS AirCal front panel. On the back right side is a small cover. Using the Allen key (supplied with the irrigator
cradle) unscrew the cover and remove.
4. Using the Allen key unscrew the internal cover and remove.
5. Expand the funnel supplied with the ICS AirCal.
6. Place the funnel in the refill hole of the reservoir.
7. Pour water into the funnel to fill the reservoir.
8. Monitor the water level closely. Fill to the bottom of the lip. Do NOT overfill the reservoir.
9. If water was spilled into the small indentions where the internal cover is placed. Wipe up this excess water before
replacing the internal cover.
10. Replace the internal cover and tighten the screws.
11. Replace the external cover and tighten the screws.
12. Plug in the ICS AirCal to the electrical wall outlet and turn on ICS AirCal.
Caution
•
Due to the possibility of freezing during shipment, water must be drained from the reservoir before shipping
the ICS AirCal. See
for water removal instructions..