Otometrics
103
App. 3 Technical Specifications
Measurement techniques
App. 3.3.14
Transport and storage environment
App. 3.3.15
Operating environment
App. 3.3.16
Warm-up time
App. 3.3.17
Disposal
MADSEN AccuScreen can be disposed of as normal electronic waste, according to
WEEE and local regulations.
App. 3.3.18
Standards
Temperature range:
-20 - +60°C (-4 - 140 °F)
Humidity range:
20-80 % rel., non-condensing
Air pressure
500 hPa to 1060 hPa
Mode of operation:
Continuous
Temperature range:
10 - 40°C (50 - 104°F)
Humidity range:
30-80 % rel., non-condensing
Air pressure
600 hPa to 1060 hPa
< 20 seconds.
Note ·
Should be extended if MADSEN AccuScreen has been stored in a cold environ-
ment.
Otoacoustic emissions:
EN 60645-6, Type 2
Auditory evoked potentials: EN 60645-7, Type 2
Patient Safety:
• EN 60601-1, Internally Powered, Type BF, IPXO
• U2601-1; CAN/CSA-C22.2 NO 601.1-90,
• IEC 60601-2-26
• IEC 60601-2-40
EMC:
EN 60601-1-2