that the probe tip and the ear canal are not blocked by cerumen or vernix. Clean
the probe, change the eartip, and then repeat the test.
• It is recommended that you repeat the test if the test conditions were not opti-
mal (too much noise, too restless a patient, etc.). If required, use a sound cabin
or a room, where the ambient noise is as low as possible.
Artifact and Stimulus values
If the
Artifact
value (artifact reject) is greater than 20%, this may indicate a noisy test.
Try to improve test conditions and repeat the test.
If the
Stimulus
value (stimulus stability) is less than 80%, the probe may have shifted
or may not be in a position to record the response. Try to reposition the probe and
repeat the test.
Note ·
Retro-cochlear hearing loss cannot be detected by TEOAE testing.
Incomplete test
5.2.4
Options in the test result screens
If the test was stopped, an error message and
detailed results are shown.
The result screen appears differently if the test
failed or was aborted during calibration.
Test result options
Comments
If you wish to add comments to the test results, press the
Comments
button. This will take you to the
Comments
screen, where you can select predefined comments or add
your own comments. See
5.3 Adding comments to a
test
X
44
.
Print results
See
7 Printing
X
61
.
Содержание AccuScreen ABR
Страница 1: ...AccuScreen OAE ABR Screener User Manual Doc no 7 50 0920 US 02 Part no 7 50 09200 US 0459 ...
Страница 8: ...8 Otometrics Table of Contents AccuScreen OAE ABR Screener User Manual ...
Страница 66: ...66 Otometrics Printing Printouts from label printer ...
Страница 82: ...82 Otometrics Handling Data in AccuScreen Deleting data ...
Страница 88: ...88 Otometrics Maintenance and Cleaning Battery safety and maintenance ...
Страница 98: ...98 Otometrics App 2 Status and Error Messages Usage and test related messages ...