Quality Tests
From the
Quality Tests
menu you can test the probe and cables connected to Accu-
Screen.
We recommend that you test the probe and cables at the beginning of each work day.
Note ·
You should always comply with local regulations and routines.
8.1
Performing Quality Tests
8.2
Probe test
Every day before you start testing on patients, you should test the probe to make sure
that it functions correctly.
Caution ·
Make sure that the probe tip has been cleaned and disinfected before you
insert it in the test cavity. See
11.4.1 Cleaning and disinfecting the test cavity
X
85
.
You can start the probe test in two ways:
1. Press the
Quality Tests
button on the
Home
screen.
2. From the
Quality Tests
menu you can access
the quality test types available in AccuScreen:
•
Probe
8.2 Probe test
X
67
•
Electrode cable
8.3 Electrode cable test
X
69
•
Ear coupler cable
8.4 Ear coupler cable test
X
70
•
Combined
(Electrode cable - Ear coupler
cable)
8.5 Combined (Electrode Cable - Ear Cou-
pler Cable)
X
71
Содержание AccuScreen ABR
Страница 1: ...AccuScreen OAE ABR Screener User Manual Doc no 7 50 0920 US 02 Part no 7 50 09200 US 0459 ...
Страница 8: ...8 Otometrics Table of Contents AccuScreen OAE ABR Screener User Manual ...
Страница 66: ...66 Otometrics Printing Printouts from label printer ...
Страница 82: ...82 Otometrics Handling Data in AccuScreen Deleting data ...
Страница 88: ...88 Otometrics Maintenance and Cleaning Battery safety and maintenance ...
Страница 98: ...98 Otometrics App 2 Status and Error Messages Usage and test related messages ...